Item Description
Segment Description
Comment Line. 20 characters can be entered.
Start Time
Start time of segment in minutes. 0.00 to 649.99 min.
End Time
Time segment will be terminated. 0.10 to 650 min.
Low Mass
Lower end of mass range to be acquired. 10 to 650 m/z.
High Mass
Upper end of mass range to be acquired. 10 to 650 m/z.
Ionization Mode
Method of ion formation in trap.
Ion Preparation
Options applied after ionization to alter the ions stored in
the trap before analysis.
Special Applications
To specify data acquisition in Profile mode (vs. Centroid)
or to specify a pre-acquisition flow sampling period.
The flow sampling segment
Item Description
Method Start Time
The time to flush the system before sample collection
begins plus the actual sampling time. Time of -30 to 0
minutes may be entered.
Flow-Sampling Start Time/
End Time
The duration of the actual sampling, from -10.00 to 0
minutes. If time = 0, no flow sampling is done.
Sample Flow Rate
Set from 1 to 100 mL/min. Must be equal to value set
on EMFC in manual mode. If EMFC is set to auto
mode, the sample flow rate specified will be
downloaded at the start of the injection
Zobrazit stránku 171