Editing Methods from the Saturn Status and Control
Window
The Saturn System Control or Acquisition Window contains a "Method:" button
that allows you to directly edit the active Method.
When you click on this button, Method Builder is opened with the active Method
loaded. You can then edit that section or any other section of the Method as
described previously.
Once you have edited the Method and closed the Method Builder window, you
are prompted to reactivate the Method. Reactivating the Method downloads the
changes to the Saturn system.
NOTE: If you access the active method from the method editor via the Star Bar,
when the method editor is closed or the file saved, the method must be
reactivated in system control. To reactivated the current method, click on the
button from the System Control toolbar.
Changing Method End Time from the Saturn Control and
Status Window
During a Saturn acquisition, the end time of the acquisition can be changed from
the Control and Status window.