331
Field Description
2nd Sample Name
(Dual Mode operation only)
Up to 19 characters
Sets the name of the second sample injected. This
column is only displayed when the 2
nd
Injection is
“Advance” or “Clean & Adv”.
2nd Inject Notes
(Dual Mode operation only)
up to 180 characters
Opens the Notes window for the selected second
sample to edit or create a note about the sample.
This column is only displayed when the 2
nd
Injection
is “Advance” or “Clean & Adv”.
Vial 0 through 99 for 8400 AutoSampler
1 through 21 for 8410 AutoInjector
Sets the AutoSampler vial number of each sample in
the SampleList. When the 2nd Injection is “Advance”
or “Clean & Adv”, two sequentially numbered vials
will be shown. The first vial will be injected on the first
injection and the next vial location will be sampled for
the second injection.
Injection Volume
0.1 to 10 µL
Sets the injection volume of the sample in microliters.
When the SampleList is edited in System Control, the
maximum value allowed is set to the syringe size
currently configured in the 3800 GC setup.
During operation, if the sample volume plus solvent
volume, internal standard addition volume, and air
gaps volume specified in the active method on the
3800 exceed the syringe volume, a 3800 hardware
fault will occur. This will cause the Star Workstation
to halt automation and reset the 3800. After changing
the method settings and/or the injection volume to
reduce the total below the installed syringe volume
shown in the 3800 setup screen, you should begin
the SampleList at the selected line to bypass the
samples that have already been run.
2nd Inj Volume
(Duplicate Mode or Dual
Mode operation only)
0.1 to 10 µL
Sets the injection volume in microliters for the second
injection of the sample in Duplicate Mode or the
injection of the second sample in Dual Mode. When
the SampleList is edited in System Control, the
maximum value allowed is set to the syringe size
currently configured in the 3800 GC setup.
During operation, if the sample volume plus solvent
volume, internal standard addition volume, and air
gaps volume specified in the active method on the
3800 exceed the syringe volume, a 3800 hardware
fault will occur. This will cause the Star Workstation
to halt automation and reset the 3800. After changing
the method settings and/or the injection volume to
reduce the total below the installed syringe volume
shown in the 3800 setup screen, you should begin
the SampleList at the selected line to bypass the
samples that have already been run.
Injectors Used
Pos 1, Pos 2, 1 then 2, or 2 then 1
Selects the 8400 Injection position to be used for the
sample. If the 8400 is not configured for both