242
Item Description
AutoSampler Type 8400, 8410, or None
Specifies whether or not an 8400 AutoSampler or 8410
AutoInjector is installed on the 3800 GC. If an 8400 or 8410
is not installed on the GC or is turned off, select None. When
8400 or 8410 is selected, the following items are displayed
for editing.
Syringe Size (uL) 5 uL or 10 uL
Selects the size of the syringe that is installed on the
8400/8410 for use with this method. A 5 microliter or a 10
microliter syringe may be used with the 8400/8410.
Injection Mode Std Split/Splitless, Std On Column, Neat, Viscous, Volatile, or
User Defined
Sets the 8400/8410 parameters for the predefined modes of
operation.
Sample Depth (%) 0 to 100 %
Specifies how far the syringe needle is to go down into the
sample vial. 100% is the bottom of the vial. 0% is the bottom
of the vial septum.
Solvent Depth (%) 0 to 100 %
Specifies how far the syringe needle is to go down into the
solvent vial. 100% is the bottom of the vial. 0% is the bottom
of the vial septum.
Default Clean Section
Default Cleaning occurs when the Stop button on the front panel of the 3800 GC,
the Reset button on the 3800 Status Window in Star Workstation System
Control, or the Stop Automation menu item in the Star Workstation System
Control Automation menu is pressed after the 8400/8410 has started its run and
before it has finished its sampling and post-injection washing operations. Default
Cleaning also occurs when the 3800 GC has a fatal automation fault after the
8400/8410 has started its run and before it has finished its sampling and post-
injection washing operations.